2- 32 Keysight B2961A/B2962A SCPI Command Reference, Edition 6
Subsystem Command Summary
Setting Source Output and Measurement
:SENS[c]:WAIT:GAIN gain
:SENS[c]:WAIT:GAIN? [DEFault|
MINimum|MAXimum]
Sets the gain value used for calculating
the measurement wait time for the
specified channel.
gain=MINimum|MAXimum|DEFault|
0
to 100
1
:SENS[c]:WAIT:OFFS offset
:SENS[c]:WAIT:OFFS? [DEFault|
MINimum|MAXimum]
Sets the offset value used for calculating
the measurement wait time for the
specified channel.
gain=MINimum|MAXimum|DEFault|
0
to 1 seconds
0
:SENS[c]:WAIT[:STAT] mode
:SENS[c]:WAIT[:STAT]?
Enables or disables the measurement wait
time for the specified channel. The wait
time is defined as the time the
measurement channel cannot start
measurement after the start of a DC
output or the trailing edge of a pulse.
mode=0 or OFF disables the
measurement wait time. The wait time is
set to 0.
mode=1 or ON enables the measurement
wait time given by the following formula.
•:SENS[c]:WAIT:AUTO ON|1
condition:
wait time = gain × initial wait time +
offset
•:SENS[c]:WAIT:AUTO OFF|0
condition:
wait time = offset
The initial wait time is automatically set
by the instrument and cannot be changed.
ON
Command Summary Reset setting

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals