EasyManua.ls Logo

Keysight N1911A

Keysight N1911A
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Power Meter Functions 2
Keysight N1911A/1912A User’s Guide 111
Average mode This feature is applicable for 8480 Series, N8480 Series,
E-Series E4410, E-Series E9300, E-Series E9320 or P-Series N1920 sensor when
connected. Below are examples of SCPI commands when used in average mode.
Send the SCPI commands below to meter using remote interface.
SENS:DET:FUNC AVER
TRIG:SOUR EXT
SENS:BUFF:COUN 2048
Optional measurement configurations:
OUTP:TRIG ON
SENS:MRAT DOUB
SENS:AVER:COUN 128
INIT:CONT ON
–*OPC
*ESR?
FETC?

Table of Contents

Other manuals for Keysight N1911A

Related product manuals