5 Using E9300 E-Series Power Sensors
196 Keysight N1911A/1912A User’s Guide
5 press to display the Meas Avg Count pop-up.
6 Use the numeric keys to enter the required value.
7 Press to complete the entry.
Procedure
Switch off step detection as follows:
1 Press
. On dual channel meters, select the required channel.
2 Use the and keys to select the Step Detect setting field.
3 Press to check the step detection to Off.
4 Press key to close the Channel Setup screen.
Achieving stable results with GSM signals
Signals with a pulse repetition frequency (PRF) close to a multiple or sub-multiple
of the 440 Hz chopper-amplifier signal generate a beat note at a frequency
between the PRF and 440 Hz. Control over the filter settings is again required to
obtain stable results.
Enter
NOTE
You should also ensure the filter is not reset when a step increase or decrease in
power is detected by switching the step detection off.
TIP The PRF of a GSM signal is approximately 217 Hz and thus requires
more averaging than most other TDMA signals. To achieve a stable
measurement use the filter setting procedures to set the .
Experimentally, a setting of 148 gives optimum results
although settings in the order of 31 or 32 give acceptable results if a
faster measurement is required.