7 Using 8480 Series Power Sensors
218 Keysight N1911A/1912A User’s Guide
Procedure
First select the table for the sensor you are using as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and press
to highlight .
Figure 7-6 Sensor table selected
3 Press to complete the selection of the calibration factor table.
4 Press again to display the measurement screen. Figure 7-7 shows which
offset table is selected.
Tables
Sensor Cal Tables
Table
On
NOTE
When no data is contained in the highlighted table, the key is disabled
(grayed out).
Table
Done
Done