Using 8480 Series Power Sensors 7
Keysight N1911A/1912A User’s Guide 221
Editing/Generating sensor calibration tables
To help achieve the best accuracy in your measurement you can enter the values
supplied for the sensors you are using by editing the installed sensor calibration
tables or by generating your own custom tables.
You cannot delete any of the 20 existing calibration tables or create any additional
tables. However, you can edit or delete the content of each table. If you need
another table you should edit and re-name one of the tables.Each calibration
table can contain a maximum of 80 frequency points.
To view the calibration tables currently stored in the power meter, press ,
, . The Sensor Tbls screen is displayed as shown in
Figure 7-6.
Table 7-2 Installed power sensor models
Table Sensor Model Table Sensor Model
0
DEFAULT
[a]
[a] DEFAULT is a sensor calibration table where the reference calibration factor and calibration factors are 100%. This
sensor calibration table can be used during the performance testing of the power meter.
5 8485A
1 8481A 6 R8486A
2
8482A
[b]
[b] The Keysight 8482B and Keysight 8482H power sensors use the same data as the Keysight 8482A.
7 Q8486A
3 8483A 8 R8486D
4 8481D 9 8487A