2 General Power Meter Functions
142 Keysight N1911A/1912A User’s Guide
During calibration, the wait symbol is displayed.
Figure 2-73 Calibration wait pop-up
Offset and relative settings are ignored during calibration.
Manual calibration
To manually calibrate the power meter and sensor combination:
1 Press and the channel softkey to start the calibration routine.
The Calibrating pop-up is then displayed.
Press and . (For dual channel meters, press ,
or as required).
For N1912A, you calibrate each channel independently by pressing the
softkey by choosing or .
TIP You can reduce the steps required to carry out the manual zero and
calibration procedure as follows: