Maintenance 10
Keysight N1911A/1912A User’s Guide 263
Instrument Self Test
If is selected, the following tests are run: (These are the same tests
which are run using the *TST? command.)
– Test Point Voltages
– Calibrator
–Fan
–Battery
– Channel Peak Path
– Channel CW Path
– Channel DAP Check
As each test takes place, the name of the test is listed on the screen. While a test
is running, the message Testing... appears beside the name of the test. As each
stage of the test is completed, the Testing... message is replaced by either
Passed or Failed.
Figure 10-1 Self test complete
When the test is complete, the result is displayed. Press to return to the
Service menu.
If the self test failed, information about the failure is displayed on the screen.