Using E9300 E-Series Power Sensors 5
Keysight N1911A/1912A User’s Guide 187
Power Meter Configuration
The P-Series power meters automatically recognize an E-Series E9300 power
sensor when it is connected. The sensor calibration data is automatically read by
the power meter. The power meter also configures the auto-averaging settings
shown in Figure 5-1 to suit the power sensor characteristics.
Figure 5-1 E9300 E-Series auto-averaging settings
-4 dBm
-10 dBm
Sensor Power
Minimum
Sensor Dynamic Range
10 dBm
2 dBm
1234
1114
Resolution Setting
Number of Averages
11416
11832
1 4 16 128
1 16 64 128
Maximum
-40 dBm
-50 dBm
-20 dBm
-30 dBm
1114
11216
121664
4 16 128 256
32 64 256 256
Upper Range
Lower Range
Sensor Power
6 dBm
0 dBm
20 dBm
12 dBm
-30 dBm
-40 dBm
-10 dBm
-20 dBm
26 dBm
20 dBm
40 dBm
32 dBm
-10 dBm
-20 dBm
10 dBm
0 dBm
E9300/1B
E9300/1H
E9300/1/4A
NOTE
These values are valid only for the power meter channel connected with E-Series
E9300 power sensor. You can also configure the settings manually – refer to
Achieving stable results with TDMA signals” on page 195 if required.