General Power Meter Functions 2
Keysight N1911A/1912A User’s Guide 141
Automatic Zero
The automatic zero can be enabled as follows:
Press ,
and toggle the to On.
The Zeroing pop-up is displayed when a zero occurs.
Manual Zeroing
To manually zero the power meter and sensor:
Press and the channel softkey.
The Zeroing pop-up is displayed.
Calibration
Calibration sets the gain of each power meter channel and P-Series sensor
combination. This is achieved without the need to connect it to the 1.0 mW Power
Reference. The power meter can be set to automatically calibrate on-the-fly while
you are using it. Alternatively, you can perform the calibration manually. The
reference calibration factor is automatically set for all P-Series power sensors.
1 of 2
Auto Zero A
NOTE
When you are performing measurements, this may cause delays in obtaining
results. During this time, it is recommended to disable the automatic zeroing.
Zero
NOTE
On dual channel meters you can zero each channel independently or both
channels sequentially by pressing the softkey - choosing , or
Zero
Zero A
Zero B
Zero Both