7 Using 8480 Series Power Sensors
214 Keysight N1911A/1912A User’s Guide
Figure 7-3 Reference Calibration Factor pop-up window
5 Press
to complete the entry.
Now zero and calibrate the power meter/sensor combination as follows:
6 Press and the channel softkey. The Zeroing pop-up is
displayed.
7 Connect the power sensor to the POWER REF output.
8 Press and the channel softkey to start the calibration routine.
The Calibrating pop-up is then displayed.
Now set the sensor calibration factor for the frequency of the signal you want to
measure.
9 Check the current reference calibration factor setting by pressing ,
. The value is displayed on the Cal Fac field.
You can reduce the steps required to carry out the zero and calibration
procedure as follows:
1 Connect the power sensor to the POWER REF output.
2 Press and . (For dual channel meters, press
, or as required).