Using 8480 Series Power Sensors 7
Keysight N1911A/1912A User’s Guide 211
R8486A
Q8486A
V8486A
W8486A
E8486A
R8486D
Q8486D
These waveguide power sensors have two connectors. Use the N-Type connector to
calibrate the power meter.
8481B
8482B
These power sensors are configured with an attenuator. Prior to calibration this
attenuator must be removed. Replace the attenuator before making measurements.
8485A
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type adapter (08485-60005) to
connect to the POWER REF.Remove this adapter before making measurements.
8485D
Prior to calibration, a Keysight 11708A 30 dB reference attenuator and an APC 3.5 (f) to
50 Ω (m) N-Type adapter (08485-60005) should be connected between the power
sensor and the POWER REF. Remove this attenuator and adaptor before making
measurements.
8487A
This sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter (08487-60001) to
connect to the POWER REF. Remove this adapter before making measurements.
8487D
Prior to calibration, a Keysight 11708A 30 dB reference attenuator and an APC 2.4 (f) to
50 Ω (m) N-Type adapter (08487-60001) should be connected between the power
sensor and the Power Ref. Remove this adapter before making measurements.
Table 7-1 8480 Series connection requirements
Sensor Connection Requirements