EasyManua.ls Logo

Keysight N1911A

Keysight N1911A
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Using 8480 Series Power Sensors
224 Keysight N1911A/1912A User’s Guide
Enter (or edit) the frequency and cal factor pairs as follows:
1 Press to add a new frequency value (or press to edit). Use the
numeric keypad to enter the required value in the Frequency pop-up window.
Complete the entry by pressing the , keys.
2 Enter the new cal factor value (or press to edit). Use the numeric
keypad to enter the required value in the Cal Factor pop-up window. Complete
the entry by pressing the key.
3 Continue adding/editing values until you have entered all the data you require.
4 When you have finished editing the table press to save the table.
NOTE
A calibration factor in the range of 1 % to 150 % can be entered.
The following rules apply to naming sensor calibration tables:
The name must consist of no more than 12 characters.
All characters must be upper or lower case alphabetic characters, or numeric
(0-9), or an underscore (_).
No other characters are allowed.
No spaces are allowed in the name.
Insert
Change
GHz
MHz
Change
%
Done
NOTE
Ensure that the frequency points you use cover the frequency range of the
signals you want to measure. If you measure a signal with a frequency outside
the frequency range defined in the sensor calibration table, the power meter
uses the highest or lowest frequency point in the sensor calibration table to
calculate the offset.

Table of Contents

Other manuals for Keysight N1911A

Related product manuals