Using N8480 Series Power Sensors 8
Keysight N1911A/1912A User’s Guide 241
– Use the and keys to highlight the Cal Fac value field and press
to display the Cal Factor pop-up. Use the numeric keypad to enter 97.6 in the
Cal Factor pop-up window.
– Press
to complete the entry.
– Connect the power sensor to the signal to be measured.
– The corrected measurement result is displayed.
%
NOTE
When no sensor tables are selected and display mode is
chosen, the calibration factor used for the measurement is displayed in the
upper window as shown in Figure 8-7.