General Power Meter Functions 2
Keysight N1911A/1912A User’s Guide 77
Enter (or edit) the frequency and offset pairs as follows:
1 Press to add a new frequency value (or press to edit). Use the
numeric keypad to enter the required value in the Frequency pop-up window.
Complete the entry by pressing the , keys.
2 Enter the new offset value (or press to edit). Use the numeric keypad
to enter the required value in the Offset pop-up window. Complete the entry by
pressing the key.
3 Continue adding/editing values until you have entered all the data you require.
4 When you have finished editing the table press to save the table.
Insert
Change
GHz
MHz
Change
%
Done
NOTE
If you measure a signal with a frequency outside the frequency range defined in
the frequency dependent offset table, the power meter uses the highest or
lowest frequency point in the frequency dependent offset table to calculate the
offset.