EasyManuals Logo

Keysight N9038A Service Guide

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
106 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–130 Suffix error A problem was found in a suffix (units). The exact problem cannot be
specifically identified
–131 Invalid suffix There is a syntax problem with the suffix. You need to use the suffix (units) that
are allowed by this command
–134 Suffix too long The suffix contained more than twelve characters
–138 Suffix not allowed A suffix was found after a numeric element that does not allow suffixes (units)
–140 Character data error A problem was found with a character data element. The exact problem cannot
be specifically identified
–141 Invalid character
data
Either the character data element contains an invalid character or the element
itself is not valid for this command
–144 Character data too
long
The character data element contains more than twelve characters
–148 Character data not
allowed
A character data element that you sent is valid, but it is not allowed at this point
in the parsing
–150 String data error A problem was found with a string data element. The exact problem cannot be
specifically identified
–151 Invalid string data A string type of data element was expected, but it is invalid for some reason.
For example, an END message was received before the terminal quote
character
–158 String data not
allowed
A string data element that you sent is valid, but it is not allowed at this point in
the parsing
–160 Block data error A problem was found with a block data element. The exact problem cannot be
specifically identified
–161 Invalid block data A block data element was expected, but it was invalid. For example, an END
message was received before the end length was satisfied
–168 Block data not
allowed
A legal block data element was found, but it is not allowed at this point in the
parsing
–170 Expression error A problem was found with an expression data element. The exact problem
cannot be specifically identified
–171 Invalid expression An expression data element is not valid. For example, there may be unmatched
parentheses or an illegal character
–178 Expression data not
allowed
A legal expression data was found, but it is not allowed at this point in the
parsing
–180 Macro error A problem was found with a macro element. The exact problem cannot be
specifically identified
–181 Invalid outside macro
definition
Indicates that a macro parameter placeholder was encountered outside of a
macro definition
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals