Keysight N9038A MXE Service Guide 79
Instrument Messages
Event Messages
–300 to –399, Device-Specific Errors
Err# Message Verbose/Correction Information
–300 Device-specific error An instrument error occurred and the exact problem cannot be specifically
identified. Report this error to the nearest Keysight Technologies sales or
service office
–310 System error; An internal system-type error has occurred. The exact problem cannot be
specifically identified. Report this error to the nearest Keysight Technologies
sales or service office
–310 System error; A
license will soon
expire;<feature code>
will expire in <time>
The indicated feature/software will expire in the specified time. Contact
Keysight Technologies to purchase continued use of this functionality
–310 System Error; enable
GPIB controller mode
Press
System, I/O Config, GPIB and set GPIB Controller to Enabled so
that the analyzer can control the source over GPIB
–310 System error; Error
transmitting a LAN
event to the network.
Communication with the network driver failed
–310 System error; Failed
to initialize the PTP
clock to current
time.
Failure communicating with the DMC libraries’ PTP controller
–310 System error; Failed
to instantiate the
PTP ordinary clock.
Failure in the starting up the DMC libraries’ PTP controller
–310 System error; feature
<feature code> not
licensed
The specified feature, for example “N9073A-TR2”, is not licensed. The license
may have expired. You cannot use it until you obtain a license
–310 System error; Feature
expired; <feature
code>
The specified feature has expired. The license is no longer valid
–310 System error; License
installation
failed;<feature code>
The license installation of the specified feature, for example "N9073A-TR2",
has failed. You should refer to the event log in the control panel for more details
–310 System error; License
removal failed;
<feature code>
The license removal of the specified feature, for example “N9073A-TR2” has
failed. You should refer to the event log in the control panel for more details
–310 System error; No
license; <feature
code> will terminate
in <time>
The specified feature will stop working in the specified time due to the license
expiration You will be prompted to save results and exit

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals