Keysight N9038A MXE Service Guide 79
Instrument Messages
Event Messages
–300 to –399, Device-Specific Errors
Err# Message Verbose/Correction Information
–300 Device-specific error An instrument error occurred and the exact problem cannot be specifically
identified. Report this error to the nearest Keysight Technologies sales or
service office
–310 System error; An internal system-type error has occurred. The exact problem cannot be
specifically identified. Report this error to the nearest Keysight Technologies
sales or service office
–310 System error; A
license will soon
expire;<feature code>
will expire in <time>
The indicated feature/software will expire in the specified time. Contact
Keysight Technologies to purchase continued use of this functionality
–310 System Error; enable
GPIB controller mode
Press
System, I/O Config, GPIB and set GPIB Controller to Enabled so
that the analyzer can control the source over GPIB
–310 System error; Error
transmitting a LAN
event to the network.
Communication with the network driver failed
–310 System error; Failed
to initialize the PTP
clock to current
time.
Failure communicating with the DMC libraries’ PTP controller
–310 System error; Failed
to instantiate the
PTP ordinary clock.
Failure in the starting up the DMC libraries’ PTP controller
–310 System error; feature
<feature code> not
licensed
The specified feature, for example “N9073A-TR2”, is not licensed. The license
may have expired. You cannot use it until you obtain a license
–310 System error; Feature
expired; <feature
code>
The specified feature has expired. The license is no longer valid
–310 System error; License
installation
failed;<feature code>
The license installation of the specified feature, for example "N9073A-TR2",
has failed. You should refer to the event log in the control panel for more details
–310 System error; License
removal failed;
<feature code>
The license removal of the specified feature, for example “N9073A-TR2” has
failed. You should refer to the event log in the control panel for more details
–310 System error; No
license; <feature
code> will terminate
in <time>
The specified feature will stop working in the specified time due to the license
expiration You will be prompted to save results and exit