Keysight N9038A MXE Service Guide 53
Boot Up and Initialization
Potential Problems During Boot Process
Instrument Hangs at the Keysight Splash Screen
A problem of the instrument hanging at the Keysight splash screen could be
caused by many different things. It could be due to a down power supply, a
processor hardware problem, an instrument boot-up process error, etc.
1. Remove the AC power cord and then remove the instrument cover. Refer
toChapter 18, “Assembly Replacement Procedures”, on page 527 in this
manual.
2. Remove the Top Brace. Refer to Chapter 18, “Assembly Replacement
Procedures”, on page 527 in this manual.
3. With the AC power applied and the On/Off button turned on, verify that all
of the power supply voltages are at their proper level. This can easily be
done by viewing the power supply LEDs on the back side of the A7
Midplane board. See the Chapter 12, “Power Supply & Midplane”, on
page 367.
Are all of the power supply voltages at the proper level?
If yes:
After verifying that the connections from the A7 Midplane
board to the A4 Processor board are not at fault, replace
the A4 Processor board.
If not:
After verifying that the connections between the A6 Power
Supply, the A7 Midplane board, and the A8 Motherboard
are all mechanically and electrically secure, replace the A6
Power Supply assembly.

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals