582 Keysight N9038A MXE Service Guide
Assembly Replacement Procedures
Conducted Filter Assembly
Conducted Filter Assembly
Removal
1. Remove the instrument outer case. Refer to the Instrument Outer Case
removal procedure.
2. Remove the top brace. Refer to the Top Brace and Power Supply Bracket
removal procedure.
3. Refer to Figure 18-39. Locate the Conducted Filter assembly (5).
4. Refer to Figure 18-42. Remove cables W32 and W33 using the 5/16-inch
wrench.
5. The Conducted Filter assembly can now be unplugged from the
motherboard by using the ejector on the Conducted Filter assembly to
disconnect from the motherboard and lift out of the chassis.
Replacement
1. Slide the Conducted Filter assembly into the chassis and press down to
plug it into the motherboard using the ejector to push into place.
2. Refer to Figure 18-42. Replace cables W32 and W33. Torque to
10 inch-pounds.
3. Replace the top brace. Refer to the Top Brace and Power Supply Bracket
replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.