Keysight N9038A MXE Service Guide 95
Instrument Messages
Event Messages
–200 to –299, Execution Errors
This section lists all messages in the range except –221 error messages. For
details of –221 messages, see “–221 Settings Conflict Errors” on page 82.
Note that Execution Errors are divided into subclasses, as follows:
–21x Trigger errors
–22x Parameter error
–23x Data corrupt or stale (invalid data)
–24x Hardware error
–25x Mass storage error
–26x Expression data error
–27x Macro error
–28x Program error (a downloaded program-related execution error)
–29x Memory use error
Err# Message Verbose/Correction Information
–200 All ranges are off.
Turn on at least a
range
There are no range turn on in scan table. You need to turn on at least a range to
initiate a scan
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom
–200 Execution Error A program execution error has occurred. The exact problem cannot be
specifically identified
–200 Execution error;
Carrier frequency
outside device’s
transmit band
The entered channel/carrier frequency is not within the range of your current
mode setup selection of standard and device
–200 Execution error;
Invalid GSM burst
timing
A GSM-like burst was acquired, but its timing is not valid. Ensure the correct
Burst Type has been selected
–200 Execution error;
Invalid IP address
The IP address supplied is either not valid or does not belong to a compatible
Signal Generator. Please check the IP address and instrument connection and
try again
–200 Execution error;
Invalid Marker Trace.
Cannot place markers on the reference trace, because the reference trace is
currently turned off or has no data
–200 Execution Error; No
peak found.
No signal peak was found within the defined parameters of the search. (Note:
for ESA/PSA compatibility, the Swept SA measurement uses 780 for this error
number; all other measurements use –220)