96 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–200 Execution error; No
ranges are defined.
Activate a range.
There are no active ranges in the range table. You will need to activate at least
one range
–200 Execution Error;
Preselector centering
failed
Algorithm failed to center the preselector. This maybe caused by the signal
peak being too low in amplitude, or it could be due to excessive CW input
signal, alignment error, or hardware failure
–200 Execution Error;
Signal not stable
enough to track
The signal that you have selected to track is changing too much for the function
to track it properly
–200 Execution Error;
Store ref trace
before turning on
Normalize
The Reference trace data must be stored in the Ref trace before you turn on the
Normalization function
–200 Execution error; Sync
word was not found.
NADC & PDC: In an EVM measurement, the sync word is not found and the
synchronization cannot be established when Sync Word is selected in the Burst
Sync menu.
Flexible Digital Demodulation: The sync word cannot be detected because of
inappropriate parameter settings or incorrect signal
–200 Execution error;
Trace file contains
no compatible traces.
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with the
version you are running. Please check that you are running the most
up-to-date version of the personality
–200 Execution error;
Trace file created by
incompatible version
of Phase Noise App
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with the
version you are running. Please check that you are running the most
up-to-date version of the personality
–200 Function not
available before
Marker Zoom is
performed
Function not available before perform Marker Zoom
–200 Must perform Scan
before do Search
Cannot perform Search as the trace data is empty
–200 No marked signal Cannot perform the selected function because no signal was marked. You must
mark the peak of interest before selecting the function
–200 No Measure At Marker
Added to Signal List
No signal peak was added into Signal List as there is no valid measure to
Marker result. You must perform Measure at Marker before selecting Measure
At Marker --> List
–200 No Peak Added to
Signal List
No signal peak was added from the Search to Signal List as no signal peaks
were found within the defined parameters of the search criteria
–200 Signal List is Empty Cannot perform the selected function because the signal list contains no data
Err# Message Verbose/Correction Information