663
Keysight Technologies X-Series Signal Analyzers
N9038A MXE Signal Analyzer
Service Guide
21 Performance Verification and Adjustment Software
Test Software Overview
The N9038A MXE EMI Receiver can be fully tested and adjusted with the use of
the N7818A MXE EMI Receiver Calibration Application software.
To download a copy of the Performance Verification & Adjustment software as
well as find information on software licensing, visit the Keysight Calibration &
Adjustment Software web site at:
http://www.keysight.com/find/calibrationsoftware
Performance Verification Tests
Performance verification tests are tests designed to provide the highest level of
confidence that the instrument being tested conforms to published,
factory-set specifications. The tests are supplied in an automated test software
package. The automatic execution of the full set of performance tests will take
between six and eight hours to complete. Performance tests are designed to
test an instrument operating within the operational temperature range defined
by the instrument specifications.
A complete list of the performance verification tests can be found at:
http://cal.software.keysight.com/
Adjustments
Adjustments are procedures designed to reset various circuit parameters or
calculate correction values associated with certain measurements. The
adjustments are supplied in an automated test software package. The software
is designed to adjust an instrument operating within the operational
temperature range defined by the instrument specifications.
Never perform adjustments as routine maintenance. Adjustments should be
performed only after a repair or a performance test failure.
A complete list of the adjustments can be found at:
http://cal.software.keysight.com/