EasyManua.ls Logo

Keysight N9038A - Replacement

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
586 Keysight N9038A MXE Service Guide
Assembly Replacement Procedures
Wideband Analog IF and Wideband Digital IF Assemblies
Replacement
1. Slide the Wideband Analog IF assembly into slot 3 in the chassis and use
the ejectors to push into place into the motherboard.
2. Slide the Wideband Digital IF assembly into slot 5 in the chassis and use
the ejectors to push into place into the motherboard.
3. Refer to Figure 18-44. Reinstall cables W15, W23, W50, W51, and W52 to
the Wideband Analog IF assembly.
4. Reattach ribbon cable W10.
5. Replace the top brace. Refer to the Top Brace and Power Supply Bracket
replacement procedure.
6. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.