574 Keysight N9038A MXE Service Guide
Assembly Replacement Procedures
Limiter Assembly
Replacement
1. Refer to Figure 18-38. Place the Limiter Assembly into the chassis.
Replace the four screws (1) (0515-0372) using the T-10 driver. Torque to
9 inch-pounds.
2. Reattach the cables W1, W6, W26, and W27 to the Limiter Assembly.
Torque the semi-rigid cables to 10 inch-pounds.
3. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
4. Replace the instrument top brace. Refer to the Top Brace and Power
Supply Bracket replacement procedure.
5. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.