580 Keysight N9038A MXE Service Guide
Assembly Replacement Procedures
Radiated Filter Assembly
Replacement
1. Slide the Radiated Filter assembly into the chassis and press down to plug
it into the motherboard using the ejector to push into place.
2. Refer to Figure 18-42. Replace cable W34. Torque to 10 inch-pounds.
3. Replace the top brace. Refer to the Top Brace and Power Supply Bracket
replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.