Keysight N9038A MXE Service Guide 363
CPU & Disk Drive
Reloading the Instrument Application Software
Reloading the Instrument Application Software
The instrument application software contains all the required components for
the EMI Receiver application as well as all software options. If the instrument
application software has become corrupt the Windows operating system will
boot but the instrument application software will fail to start. In these cases go
to Chapter 20, “Instrument Software” for information on how to update the
instrument application software. This procedure will not affect the instrument’s
calibration or user files.
If this does not resolve the boot issue, or the instrument never boots the
Windows operating system proceed to the “Disk Drive Recovery Process”.