Keysight N9038A MXE Service Guide 605
Assembly Replacement Procedures
AIF/DIF Assembly
Replacement
1. Slide the AIF/DIF assembly into the slot at the rear of the instrument and
push on the assembly to mate the connectors to the motherboard
assembly.
2. Refer to Figure 18-54. Replace cables W16, W17, and W19 to the AIF
assembly.
3. Refer to Figure 18-53. Replace cable W15, W17, W20, and W23 to the
bottom of the DIF assembly.
4. Replace the rear panel. Refer to the Rear Panel replacement procedure.
5. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.