532 Keysight N9038A MXE Service Guide
Assembly Replacement Procedures
Instrument Outer Case
Instrument Outer Case
Removal
1. Disconnect the instrument from ac power.
2. Refer to Figure 18-3. Using the T-20 driver, remove the four screws (two
on each side) (1) that attach the handle strap (2) on each side of the
instrument.
3. Remove the bottom feet and locks (6) by lifting the tabs on the feet and
sliding to disengage from the outer case.
4. Using the T-20 driver, remove the four screws (including washers) (3) that
hold the rear feet (4) in place.
5. Pull the instrument cover (5) off towards the rear of the instrument.
If the instrument is placed on its face during any of the following procedures, be sure to use a soft
surface or soft cloth to avoid damage to the front panel, keys, or input connector.