Keysight N9038A MXE Service Guide 97
Instrument Messages
Event Messages
–200 Signal List is Full Cannot perform the selected function because the signal list is full. Please
clear the list
–200 Signal Selected is
not in the Signal
List
The signal selected is not the list. You only can perform the operation on signal
that is already available in the list
–201 Invalid while in
local
The command cannot be executed while the instrument in Local control
–202 Settings lost due to
rtl
A “return to local” control was forced and some settings were lost as a result of
this
–203 Command protected The command could not be executed because it is disabled. It was disabled by
licensing or password protection
–203 Command protected;
feature not licensed
The specified feature, for example “N9073A-TR2” is not licensed. The license
may have expired. You cannot use it until you obtain a license
–210 Trigger error A trigger error has occurred, but the exact problem cannot be specifically
identified
–211 Trigger ignored A GET, *TRG or other triggering signal was received, but was ignored because
of timing considerations. For example, maybe the instrument was not ready to
respond when the command was received
–212 Arm ignored An arming signal was received, but it was ignored
–213 Init ignored An initiate trigger/sweep request was received and ignored, because another
measurement was already in progress
–214 Trigger deadlock The trigger source for the initiation of a measurement is set to GET, and the
following measurement query was received. The measurement cannot be
started until a GET is received, but the GET would cause an INTERRUPTED
error
–215 Arm deadlock The arm source for the initiation of a measurement is set to GET and the
following measurement query is received. The measurement cannot be started
until a GET is received and the GET would cause an INTERRUPTED error
–220 Parameter error A problem was found with a program data element. The exact problem cannot
be specifically identified
–221 Settings conflict;
There are many types of settings conflict errors. For full details, see
“–221
Settings Conflict Errors” on page 82
–222 Data out of range; A data element was found but the instrument could not be set to that value
because it was outside the range defined for the command. A descriptive
message may be appended, such as “clipped to upper limit”
–222 Data out of Range;
clipped to source
max/min
A source parameter has been entered that exceeds the range of the selected
source. The parameter has been clipped to match the range of the source
Err# Message Verbose/Correction Information