80 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–310 System Error; No
supported source
found
Signal source at given IP address is not responding / IP does not belong to a
source. Check IP address and network connection
–310 System Error; source
connection lost,
check interface
connection
Signal source at given IP address is not responding / IP does not belong to a
source. Check IP address and network connection
–310 System error; The
configured PTP
hardware driver could
not be instantiated.
The PTP driver failed on initialization
–310 System error; The PTP
hardware driver
reported a
configuration error.
Failure in the execution of the PTP driver. The most likely cause of this error is a
mismatch between versions of the PTP driver and the LXI middleware
–310 System error; The PTP
ordinary clock
reported a
configuration error.
Failure in execution of the DMC libraries’ PTP controller
–310 System error; The
Trigger alarm delayed
LAN event was not
scheduled due to an
existing pending
event.
Delayed LAN events cannot occur too close together (within 20 ms)
–310 System error; The
Trigger alarm delayed
LAN event was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Delayed LAN events cannot occur too close to a scheduled Alarm (within 20
ms)
–310 System error; The
Trigger alarm was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Alarms cannot be scheduled to happen too close together (within 20 ms)
–311 Memory error There is a physical problem with the instrument memory, such as a parity error
–312 PUD memory lost Protected user data saved by the *PUD command has been lost
–313 Calibration memory
lost
The nonvolatile calibration data used by the *CAL? command has been lost