Keysight N9038A MXE Service Guide 81
Instrument Messages
Event Messages
–314 Save/recall memory
lost
The nonvolatile data saved by the *SAV? command has been lost
–315 Configuration memory
lost
The nonvolatile configuration data saved by the instrument has been lost
–320 Storage fault; A problem was found while using data storage. The error is not an indication of
physical damage or failure of any mass storage element
–321 Out of memory An internal operation needed more memory than was available. Report this
error to the nearest Keysight Technologies sales or service office
–330 Self-test failed A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office
–340 Calibration failed The instrument requires an AlignAllNow. Restore the alignment by pressing
System, Alignments, AlignAllNow
–340 Calibration failed;
<failure msg>
<port>
The calibration for one of the I-Q ports did not succeed. The information in the
“failure msg” field can be used to troubleshoot this problem. Contact Keysight
technical support
–350 Queue overflow An error occurred that did not get put in the error queue because the queue
was full
–360 Communication error There was a problem with instrument remote communications. The exact
problem cannot be specifically identified
–360 Communication error;
SNS data read
failure. Disconnect
then reconnect SNS
The Keysight Smart Noise Source connected to the instrument has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or another
hardware fault exists. Check SNS on another instrument, NFA and ESA are also
SNS compatible instruments. Check that the device is not an Keysight power
sensor which uses the same cable interface
–360 Communication error;
SNS is not connected
The Keysight Smart Noise Source connected to the instrument has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or another
hardware fault exists. Check SNS on another instrument, NFA and ESA are also
SNS compatible instruments. Check that the device is not an Keysight power
sensor which uses the same cable interface
–361 Parity error in
program message
A parity bit was not correct when the data was received. For example, on a
parallel port
–362 Framing error in
program message
A stop bit was not detected when data was received. For example, on a remote
bus port
–363 Input buffer overrun A software or hardware input buffer on a port overflowed with data because of
improper or nonexistent pacing
–365 Time out error There was a time-out problem in the instrument. The exact problem cannot be
specifically identified
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals