Keysight N9038A MXE Service Guide 195
RF Preselector Section
RF Preselector Section Troubleshooting
Figure 5-20 A22 Radiated Filter Assembly LEDs
Radiated Filter Board Heartbeat
This LED should start blinking just a few seconds after the instrument power is
turned on. If it does not come on at all there is either a problem with an
instrument power supply voltage or the FPGA chip on the A22 Radiated Filter
assembly is not getting programmed properly. To verify the instrument power
supply voltages see Chapter 12, “Power Supply & Midplane”. If the instrument
power supply voltages are all correct try reload the FPGA code. See FPGA
Table 5-7 A22 Radiated Filter Assembly Status LEDs
LED LED Color Description Status
A22DS410 Green Heartbeat Blinking = Normal Operation
On = FPGA Reference Problem
Off = FPGA Programming Problem
A22DS409 Green Fixed Frequency Mode On = Enabled
Off = Disabled
A22DS408 Green Swept Mode On = Enabled
Off = Disabled
A22DS407 Green Stepped Mode On = Enabled
Off = Disabled
A22DS406 Yellow Not Used On = N/A
Off = Normal Operation
A22DS405 Yellow Not Used On = N/A
Off = Normal Operation

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals