Keysight N9038A MXE Service Guide            329
Analog & Digital IF
A2 Analog I.F. Troubleshooting
Figure 10-3  A2 Analog IF 22.5 MHz IF Output with Low Gain
12.Set the step gain to High by pressing Meas Setup, More, Swept IF Gain, 
High Gain.
13.Verify that the 22.5 MHz IF output from the A2 Analog IF assembly 
measures between -11 and -19 dBm as shown in Figure 10-4.