Keysight N9038A MXE Service Guide 619
Assembly Replacement Procedures
Input Connector Assembly
Replacement
1. Refer to Figure 18-65. Place the Input Connector Assembly into position in
the chassis. Replace the two screws (1) (0515-0372) to attach the Input
Connector Assembly to the chassis. Torque to 9 inch-pounds.
2. Reconnect the semi-rigid cable W25 to the Input Connector Assembly.
Torque to 10 inch-pounds.
3. Replace the Front Frame Assembly. Refer to the Front Frame Assembly
replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.