EasyManuals Logo

Keysight N9038A Service Guide

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #600 background imageLoading...
Page #600 background image
Keysight N9038A MXE Service Guide 619
Assembly Replacement Procedures
Input Connector Assembly
Replacement
1. Refer to Figure 18-65. Place the Input Connector Assembly into position in
the chassis. Replace the two screws (1) (0515-0372) to attach the Input
Connector Assembly to the chassis. Torque to 9 inch-pounds.
2. Reconnect the semi-rigid cable W25 to the Input Connector Assembly.
Torque to 10 inch-pounds.
3. Replace the Front Frame Assembly. Refer to the Front Frame Assembly
replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals