Keysight N9038A MXE Service Guide            621
Assembly Replacement Procedures
Input Connector Assembly
Replacement 
1. Refer to Figure 18-66. Place the Input Connector Assembly into position in 
the chassis. Replace the two screws (1) (0515-0372) to attach the Input 
Connector Assembly to the chassis. Torque to 9 inch-pounds.
2. Reconnect the semi-rigid cable W25 to the Input Connector Assembly. 
Torque to 10 inch-pounds.
3. Replace the Front Frame Assembly. Refer to the Front Frame Assembly 
replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case 
replacement procedure.