Keysight N9038A MXE Service Guide            649
Post-Repair Procedures
Post-Repair Procedures
If the Instrument Software Application Will Not Run
1. Connect a USB mouse to one of the instrument front panel USB ports.
2. Connect a USB FLASH drive to the other front panel USB port.
3. Using the mouse go to Start, Computer, and navigate to the following 
folder:  
 
C:\Program Files\Agilent\SignalAnalysis\Infrastructure
4. Double-click on BackupRestoreAlignDataWizard and the window shown in 
Figure 19-1 will appear.
5. Follow the on-screen instructions to backup the calibration data to the 
USB FLASH drive.