Keysight N9038A MXE Service Guide 669
Tests
Test Equipment
Cables
Cable, 1 meter
2.4 mm (m) to 2.4 mm (m)
1
b
Frequency:50 MHz to 44 GHz
VSWR< 1.55:1
8120-6164
Cable, 1 meter
3.5 mm (m) to 3.5 mm (m)
2 Frequency:10 MHz to 26.5 GHz
VSWR< 1.4:1
11500E
Cable, 120 cm
BNC (m) to BNC (m)
2 Frequency:10 MHz 10503A
Power Meters/Sensors
Power Meter 1 Compatible with power sensor E4418B
Power Sensor 1 Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −25 dBm to 10 dBm
8485A
Power Sensor 1 Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: –65 to –55 dBm
8485D
Power Sensor
1
b
Frequency Range: 50 MHz to 44 GHz
Amplitude Range: –30 to 20 dBm
8487A
Power Sensor
1
b
Frequency Range: 50 MHz to 44 GHz
Amplitude Range: –65 to –55 dBm
8487D
Power Sensor 1 Frequency Range: 9 kHz to 1 GHz
Amplitude Range: –60 to 0 dBm
E9304A
Signal Source
Signal Generator 1 Frequency: 50 MHz to 26.5 GHz
Output Level Accuracy: 0 to −15 dBm: ±1.0 dB
Spectral Purity: Better than −20 dBc
Typical Temperature Stability: 0.01 dBc/°C
PSG
EMI Calibration Pulse
Generator
1 Pulse Area meets CISPR 16-1-1 specification Schwarzbeck
IGUU 2916
Miscellaneous Equipment
Filter, 50 MHz Low Pass
BNC (m) to BNC (f)
1 Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
0955-0306
Power Splitter
3.5 mm (f)
1 Frequency Range: 50 MHz to 26.5 GHz
Tracking Between Ports: < 0.25 dB
11667B
Table 22-1 Required Test Equipment
Item Quantity Critical Specifications
Recommended
Keysight Model