Page #674 background imageLoading...
Page #674 background image
Keysight N9038A MXE Service Guide 693
Tests
Frequency Response (Flatness)
Table 22-17 Frequency Response Results - RF Input 1
Center
Frequency
Receiver
Amplitude
(dBm)
Meas_Amp
Power Meter
Amplitude
(dBm)
Power_Meter
Measurement
Error
(dB)
Meas_Error
Normalized
to 50 MHz
(dB)
Normalized
Test Limit
(dB)
50 MHz 0 Ref
1.0 GHz ±1.75 dB
2.0 GHz ±1.75 dB
3.0 GHz ±1.75 dB
4.0 GHz ±2.5 dB
6.0 GHz ±2.5 dB
8.0 GHz ±2.5 dB
9.0 GHz ±2.5 dB
11 GHz ±2.5 dB
13 GHz ±2.5 dB
14 GHz ±2.5 dB
17 GHz ±2.5 dB
20 GHz ±3.0 dB
23 GHz ±3.0 dB
26 GHz ±3.0 dB
29 GHz ±3.5 dB
32 GHz ±3.5 dB
35 GHz ±4.5 dB
38 GHz ±4.5 dB
41 GHz ±4.5 dB
44 GHz ±4.5 dB