EasyManua.ls Logo

Keysight N9038A - Page 674

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N9038A MXE Service Guide 693
Tests
Frequency Response (Flatness)
Table 22-17 Frequency Response Results - RF Input 1
Center
Frequency
Receiver
Amplitude
(dBm)
Meas_Amp
Power Meter
Amplitude
(dBm)
Power_Meter
Measurement
Error
(dB)
Meas_Error
Normalized
to 50 MHz
(dB)
Normalized
Test Limit
(dB)
50 MHz 0 Ref
1.0 GHz ±1.75 dB
2.0 GHz ±1.75 dB
3.0 GHz ±1.75 dB
4.0 GHz ±2.5 dB
6.0 GHz ±2.5 dB
8.0 GHz ±2.5 dB
9.0 GHz ±2.5 dB
11 GHz ±2.5 dB
13 GHz ±2.5 dB
14 GHz ±2.5 dB
17 GHz ±2.5 dB
20 GHz ±3.0 dB
23 GHz ±3.0 dB
26 GHz ±3.0 dB
29 GHz ±3.5 dB
32 GHz ±3.5 dB
35 GHz ±4.5 dB
38 GHz ±4.5 dB
41 GHz ±4.5 dB
44 GHz ±4.5 dB

Table of Contents

Other manuals for Keysight N9038A

Related product manuals