EasyManua.ls Logo

Keysight N9038A - Page 703

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
722 Keysight N9038A MXE Service Guide
Tests
Quasi-Peak Detector Accuracy
27.Change the EMI receiver input attenuation to 10 dB by pressing AMPTD Y
Scale, Attenuation, 10 dB.
28.Take a single sweep on the EMI receiver by pressing Restart.
29.Reset the delta marker by pressing Marker, Delta.
30.Change the Pulse Frequency on the IGUU 2916 EMI pulse generator to
1000 Hz.
31.Take a single sweep on the EMI receiver by pressing Restart.
32.Once the sweep on the EMI receiver has completed and the auto peak
value is found record the marker delta amplitude value for the pulse
frequency of 1000 Hz in Table 22-37.

Table of Contents

Other manuals for Keysight N9038A

Related product manuals