After you select the clinical option/preset, the system calibrates the transducer,
enables the transducer for operation, and updates system status to reflect
the transducer type and the preset you selected.
ECG Feature
The system can display a trace on the image representing an ECG signal. ECG
input can be either a low-level signal (from leads connected to the patient) or a
high-level signal (from a patient monitor). Separate receptacles are used to
connect low-level ECG and high-level ECG signals to the system. Heart rate,
derived from the ECG signal, is displayed on the screen whenever ECG is
connected and displayed.
DVD, CD, and USB Devices
Removable media compatible with the system include DVD, CD, and USB storage
devices.
You can use this drive to store and transfer patient files, including full studies,
3D data sets, and reports. Also, you can save, restore, and distribute setups data,
including presets.
For additional information on specific applications of the DVD drive, see the
Help. To display the Help, press Help on the keyboard.
Media Compatibility
CD media capacity is approximately 700 MB; DVD media capacity is
approximately 4.7 GB. Rewritable media types (indicated by the RW suffix) can
be erased and used again.
DVD and CD media are available in a number of types. Not all media types are
fully compatible with the system DVD drive. To ensure consistent performance,
use only high-quality DVD media in the system DVD drive. For more information,
see Media Compatibility on the User Information CD.
iU22 User Manual
166
4535 614 45861
Using the System
5