Manipulating the S7-2omni Tip
Review the warnings and caution in "Patient Safety During TEE Studies" on page
225 and "TEE Deflection Control Basics" on page 234 before using the transducer
in a study.
1. Turn the detent brake actuator fully counter clockwise to put both knobs
into freewheeling mode.
2. Turn the large knob to deflect the tip in the anterior/posterior plane.
3. Turn the small knob to deflect the tip in the medial/lateral plane.
4. Once the tip is positioned properly, do one of the following:
• Turn the detent brake actuator fully clockwise to put both knobs in detent
mode.
• Center the detent brake actuator to put only the small knob (medial/lateral
movement) in the detent mode.
243
iU22 User Manual
4535 614 45861
10
Transesophageal Transducers