The following cautions can help to reduce ESD effect:
CAUTIONS
• Do not touch transducer connector pins or the system’s transducer
receptacle.
• Handle the transducer by the metal connector shell.
• Make contact with a metal surface of the system before connecting a
transducer to the system.
• The following precautions can help to reduce ESD: anti-static spray on
carpets; anti-static spray on linoleum; anti-static mats; or a ground wire
connection between the system and the patient table or bed.
•
On connectors labeled with the ESD sensitivity symbol , do not touch
the connector pins, and always observe the preceding ESD precautions
when handling or connecting transducers.
NOTE
Electrostatic discharges (ESDs) may cause the ECG heart rate display to increase
by 10% to 15% for a few seconds after the discharge. However, the ECG heart
rate display will return to normal within 4 seconds.
Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified in
the table. The customer or the user of the system should ensure that it is used
in such an environment.
iU22 User Manual
70
4535 614 45861
Safety
2