Repetitive Strain References
Pike, I., et al. "Prevalence of Musculoskeletal Disorders and Related Work and
Personal Factors Among Diagnostic Medical Sonographers." Journal of Diagnostic
Medical Sonographers, Vol. 13, No. 5: 219-227, September 1997.
Necas, M. "Musculoskeletal Symptomatology and Repetitive Strain Injuries in
Diagnostic Medical Sonographer." Journal of Diagnostic Medical Sonographers,
266-227, November/December 1996.
Foot Switch Warning
WARNING
Do not use the foot switch in the operating room. IEC 60601-1 specifies that
foot-operated control devices used in the operating room must be of watertight
construction. The foot switch supplied with the ultrasound system meets only
drip-proof construction requirements.
Philips Transducers
Use only transducers that are approved by Philips for use with your Philips
ultrasound system. See "Clinical Options and Transducers" on page 196 for a list
of the transducers that are compatible with your ultrasound system.
In the United States, the FDA 510(k) regulatory clearance for use of the product
is applicable only when Philips-manufactured transducers are connected to the
system.
Glutaraldehyde Exposure
The United States Occupational Safety and Health Administration (OSHA) has
issued a regulation covering levels of acceptable glutaraldehyde exposure in the
working environment. Philips does not sell glutaraldehyde-based disinfectants
with its products, but this type of disinfectant is recommended for the disinfection
of transducers used in TEE, intraoperative, endocavity, and biopsy procedures.
65
iU22 User Manual
4535 614 45861
2
Safety