2. Disconnect the transducer from the system and remove any accessories
attached to or covering the transducer.
3. To remove all organic matter and other residue, use a pre-cleaner or detergent
to assist in removing protein residuals. Enzymatic cleaners should have a pH
of 6.0 to 8.0. Those cleaners are further diluted during use. Follow the
manufacturer’s instructions for dilution. When cleaning the lens, use a blotting
motion rather than a wiping motion.
4. To remove remaining particulate and cleaning residue, rinse thoroughly with
water up to the immersion point shown in "Disinfection of Transducers by
Immersion (High-Level Disinfection)" on page 293.
5. Wipe with a dry cloth. To dry the lens, use a soft cloth and a blotting motion
instead of a wiping motion.
Disinfecting Transducers using a Wipe or Spray Method
To disinfect transducers, you can use either an immersion method or a wipe
method with a disinfectant recommended by Philips Ultrasound. Use the method
that is biologically appropriate, as described in "Choosing a Disinfectant" on page
287.
This topic provides instructions on using the wipe or spray method.
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting any
equipment.
NOTE
Transducers can be disinfected using the wipe method only if the product labeling
of the compatible disinfectant you are using indicates it can be used with a wipe
method.
289
iU22 User Manual
4535 614 45861
13
Transducer Care