CAUTIONS
Using non-recommended disinfectants, using incorrect solution strengths,
or immersing a transducer deeper or longer than recommended can damage
or discolor the transducer and voids the transducer warranty.
Do not immerse transducers longer than the minimum time needed for
your level of disinfection. For information on the levels of disinfection
requirements, see "Choosing a Disinfectant" on page 287.
Disinfecting Transducers by Immersion
Before performing this procedure, read "Disinfection of Transducers by
Immersion (High-Level Disinfection)" on page 293.
1. Clean the transducer according to "General Cleaning for All Transducers"
on page 287.
2. Mix the disinfection solution compatible with your transducer (see
"Disinfectants Compatibility Table" on page 304) according to label instructions
for solution strength. A disinfectant listed in the table with the footnote “FDA
510(k) cleared” is recommended in the U.S.
3. Immerse the transducer into the appropriate disinfectant for your transducer
as shown. Follow the instructions on the disinfectant label for the duration
of transducer immersion. Do not immerse transducers longer than the
minimum time needed for your level of disinfection.
4. Using the instructions on the disinfectant label, rinse the transducer up to
the point of immersion, and then air dry or towel dry with a sterile cloth.
5. Examine the transducer for damage, such as cracks, splitting, fluid leaks, or
sharp edges or projections. If damage is evident, discontinue use of the
transducer and contact your Philips Ultrasound representative.
iU22 User Manual
294
4535 614 45861
Transducer Care
13