8. Lightly wipe only the steering mechanism and handle with a pad moistened
with rubbing alcohol (70% isopropyl alcohol), or use approved cleaners as
directed on the handle and steering mechanism.
9. Hang the transducer on a wall-mounted rack and let it air dry.
CAUTION
Never sterilize the transducer with autoclave, ultraviolet, gamma radiation,
gas, steam, or heat sterilization techniques. Severe damage will result.
Avoidable transducer damage is not covered by the warranty or service
contract.
Disinfecting TEE Transducers in an Automated
Disinfector
The TD-100 is an automated disinfector available from PCI Medical. For more
information about this product, see www.pcimedical.com. Use this disinfector
only for those transducers listed as compatible in "Disinfectants Compatibility
Table" on page 304 or in the results of a compatible-solution search on the
Transducer Care Web site:
www.healthcare.philips.com/us/products/ultrasound/transducers/transducer_care
1. Disconnect the transducer from the system.
2. Use the following procedure to remove all organic matter and other residue:
Use a pre-cleaner or detergent to assist in removing protein residuals.
Enzymatic cleaners should have a pH of 6.0 to 8.0. Those cleaners are
further diluted during use. Follow the manufacturer’s instructions for
dilution.
Alternatively, you may gently wipe the distal tip and flexible shaft up to the
control housing (steering mechanism) with gauze pads soaked in mild,
soapy water. Do not use iodine-based soaps.
CAUTION
Do not rinse or immerse the control housing, cable, or connector.