3. Use water to rinse the distal tip and flexible shaft thoroughly.
4. Disinfect the distal tip and flexible shaft in the TD-100 Automated TEE
Disinfector. Follow the manufacturer’s instructions for operation of the
disinfector.
5. Check the transducer for any residual organic material. If any is present,
remove it and disinfect the transducer again.
6. Gently dry the distal tip and flexible shaft with a sterile cloth or pad, or allow
it to air dry.
7. Lightly wipe only the steering mechanism and handle with a pad moistened
with rubbing alcohol (70% isopropyl alcohol), or use approved cleaners as
directed on the handle and steering mechanism.
8. Hang the transducer on a wall-mounted rack and let it air dry.
CAUTIONS
• The transducer steering mechanism is not sealed. If disinfectant or other
fluid enters the steering mechanism, it will corrode the gears and electrical
connections. Avoidable transducer damage is not covered by the warranty
or service contract.
• Never sterilize the transducer with autoclave, ultraviolet, gamma radiation,
gas, steam, or heat sterilization techniques. Severe damage will result.
Avoidable transducer damage is not covered by the warranty or service
contract.
Sterilizing Transducers
Sterilization is required if the device is classified as a critical device.
299
iU22 User Manual
4535 614 45861
13
Transducer Care