13 Transducer Care
This section contains information on cleaning, disinfecting, and sterilizing
transducers compatible with your system, as well as cleaning and disinfecting
system surfaces. This section also lists the ultrasound gels that are safe to use
with the transducers compatible with your system.
These instructions are intended to assist in effective cleaning, disinfection, and
sterilization. In addition, these instructions will help avoid damage, which could
void your warranty, during cleaning, disinfection, sterilization, and gel use.
Disinfectants and Gels Safety
Observe the following warnings and cautions when using disinfectants and gels.
More specific warnings and cautions are included within the various procedures
in this section and on the labels of the cleaning or disinfection solutions.
WARNINGS
Disinfectants listed in "Disinfectants Compatibility Table" on page 304 are
recommended because of their chemical compatibility with product
materials, not their biological effectiveness. For the biological effectiveness
of a disinfectant, see the guidelines and recommendations of the
disinfectant manufacturer, the U.S. Food and Drug Administration, and
the U.S. Centers for Disease Control.
The level of disinfection required for a device is dictated by the type of
tissue it will contact during use. Ensure the disinfectant type is appropriate
for the type of transducer and the transducer application. For information
on the levels of disinfection requirements, see "Choosing a Disinfectant"
on page 287. Also, see the disinfectant label instructions and the
recommendations of the Association for Professionals in Infection Control,
the U.S. Food and Drug Administration, and the U.S. Centers for Disease
Control.
Sterile transducer covers with sterile ultrasound transmission gel are
required for intraoperative and biopsy procedures, and protective covers
285
iU22 User Manual
4535 614 45861