Loading...
Pressure Limits
Operating: 525 mmHg to 795 mmHg (700 hPa to 1,060 hPa)
Storage: 375 mmHg to 795 mmHg (500 hPa to 1,060 hPa)
Humidity Limits
Operating: 35% to 85% (with peripherals attached); 15% to 95% (without
peripherals attached)
Storage: 15% to 95%
Temperature Limits
Operating: 10°C to 40°C (50°F to 104°F)
Storage: -34°C to 65°C (-29°F to 149°F)
Safety Requirements
Classication
Class I equipment with Type B, Type BF, and Type CF applied parts
Ordinary Equipment
Non-AP/APG
Electromechanical Safety Standards Met
CAN/CSA 22.2 No. 60601-1-08, Medical Electrical Equipment: General
Requirements for Basic Safety and Essential Performance
IEC 60601-1:2005, Medical Electrical Equipment, General Requirements for
Safety
EC 60601-1-2:2007, Collateral Standard, Electromagnetic Compatibility
IEC 60601-2-37:2007, Particular Requirements for Safety: Ultrasonic Medical
Diagnostic and Monitoring Equipment
AMI/ANSI ES60601-1:2005, Medical Electrical Equipment, General
Requirements for Safety