Procedures for using setup options and descriptions of settings are included
throughout the Help.
Changing Setups
1. Press the Setup key.
2. Click a setup category on the left side of the setups display.
3. Click a sub-category tab at the top of the setups display.
4. Enter text or make selections necessary to set up your system.
5. Click Done.
Hiding the Doppler Velocity Minus Sign
You can choose to include only the numeric value of the Doppler velocity
measurements. Doppler velocity is marked as negative (-) when the flow is
moving away from the transducer. When reporting on the measurements, you
may want only the value, not the direction of flow. When hidden, the negative
sign does not appear on the displayed or printed patient reports; however, the
sign is still part of the number and is visible when editing a patient report and is
included in DICOM structured reports exported from the system.
The option to hide the doppler velocity minus sign is available in all calculation
packages except Adult Echo.
NOTE
When Hide Doppler Velocity Minus Sign is enabled, avoid using the Doppler
velocity value displayed in the results and report for manual calculations. The
sign value, whether visible or not, affects any calculations performed with the
reported value. The system includes the sign value in calculations.
To specify the display of the doppler velocity minus sign:
a. Press the Setup key.
b. Select Analysis Config.
c. Select calculations package tab you want to change, and then select Tools &
Results.