• To avoid risk of a burn hazard, do not use transducers with high-frequency
surgical equipment. A burn hazard may result from a defect in the
high-frequency surgical neutral electrode connection.
• Using cables, transducers, and accessories other than those specified for
use with the system may result in increased emissions from, or decreased
immunity of, the system.
CAUTIONS
• Although your system has been manufactured in compliance with existing
EMI/EMC requirements, use of this system in the presence of an
electromagnetic field can cause momentary degradation of the ultrasound
image. When interference is present or intermittent, use caution when
continuing to use the system. If interference occurs often, review the
environment in which the system is being used, to identify possible sources
of radiated emissions. These emissions could be from other electrical devices
used within the same room or an adjacent room. Communication devices
such as cellular phones and pagers can cause these emissions. The existence
of radio, TV, or microwave transmission equipment located nearby can
cause emissions. In cases where EMI is causing disturbances, it may be
necessary to relocate your system.
• For information on electromagnetic emissions and immunity as it applies
to the system, see "Electromagnetic Compatibility" on page 67. Ensure that
the operating environment of your system meets the conditions specified
in the referenced information. Operating the system in an environment that
does not meet those conditions may degrade system performance.
Debrillators
Observe the following warnings when using a transducer when a defibrillation
is required.
iU22 User Manual
32
4535 614 45861
Safety
2